Empower gives "setup failure" message with Agilent LC- WKB3202

Article number: 3202

SYMPTOMS

  • Sample set fails and "Setup failure" message appears

ENVIRONMENT

  • Agilent ICF v2.1
  • Agilent ICS v1.06
  • Empower 3 Enterprise
  • Agilent 1260 Infinity

CAUSE

The system is not configured for the method to control the thermostat.

FIX or WORKAROUND

  1. Select Thermostat Controlled by Method in System Configuration.

ADDITIONAL INFORMATION