Cannot communicate with 7890 GC in Empower - WKB31513

Article number: 31513

SYMPTOMS

  • 7890 says OK? NO in node properties
  • IP address is set to static, and all information was correctly added to DHCP
  • IP address pings successfully
  • Firmware is A.01.13, which is correct for driver 7890 Instrument Control Software v2.5

ENVIRONMENT

  • Empower 3 Enterprise
  • 7890 GC
  • 7890 Instrument Control Software v2.5

CAUSE

Corrupt driver on LAC/E32

FIX or WORKAROUND

  1. Uninstall 7890 Instrument Control Software v2.5.
  2. Reboot the LAC/E32.
  3. Reinstall 7890 Instrument Control Software v2.5.
  4. Reboot the LAC/E32.
  5. Scan the instrument on the instrument tab of node properties. It should now say OK? Yes.

ADDITIONAL INFORMATION

 

id31513, 7890 GC, 7890 GCDET, 7890BGC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP