Agilent G1888 Headspace is not communicating in Empower - WKB31511
Article number: 31511
SYMPTOMS
- Agilent G1888 Headspace is not communicating in Empower; it is OK? NO in the Instrument tab of node properties
- Recent installation of HCO v2.0 driver
- IP address is static on the G1888 and pings successfully
ENVIRONMENT
- Empower 3
- Headspace control option 2.0
- Agilent G1888 headspace
CAUSE
Incorrect MAC address within DHCP for 1888
FIX or WORKAROUND
- Edit the entry in DHCP using the correct MAC address.
- Re-scan the instrument on the Instrument tab in node properties and the instrument should now appear as OK? Yes.
ADDITIONAL INFORMATION
Fixed IP addressing is required for G1888A Headspace control. Configure the communications on the G1888A HS as follows:
1. From the front panel of the G1888, press “.” , then press “4” to view the LAN Configuration screen.
2. Scroll to enter the fixed IP address. Then enter the subnet, and gateway information.
3. Press “.” , then press “5” to view the MAC Address screen.
4. Enter the IP address and MAC address in the Empower DHCP Configuration window along with Instrument Type and Unique Name.
See article WKB97285 on Headspace Sample preparation
