How to process interrupted PDA channels in Empower - WKB313404

Article number: 313404

OBJECTIVE or GOAL

Process the interrupted PDA channels in Empower (when Instrument failure occurs before the injection starts and the message "problem reading data file in review window" appears).

ENVIRONMENT

  • Empower 

PROCEDURE

  1. Create a new method set and select the related instrument method.
  2. Create a new derived channel and select the PDA MAX option.
  3. Select the scanning range.
  4. Save the method set and process using this newly created method set.

ADDITIONAL INFORMATION