Error : Data Incomplete in Empower Project review window - WKB3132
Article number: 3132
SYMPTOMS
- All injections run properly, but last injection shows "data incomplete" when brought in review
- Recover.log located in C:\Empower\InstrumentServer shows no buffering
ENVIRONMENT
- Empower 2
- Empower 3
CAUSE
Unknown
FIX or WORKAROUND
- Click on Browse Projects.
- Open the project and click Injections tab.
- Highlight the injection, right-click, and then click Review.
- Click in the chromatogram window, and then right-click.
- Click Complete Data.If you do not see Complete Data as an option, it indicates that the "Verify incomplete data in raw data files" privilege is not selected.
- To select this privilege:Click Configure the System, click User Type, highlight the user type, and click Properties. In Data Acquisition, select "Verify incomplete data in raw data files". Note: You need administrator privileges to assign this privilege.