Empower: Maurice Status - Communication with the instrument failed; Run aborted instrument already in use - WKB295959
Article number: 295959
SYMPTOMS
- Unable to launch an analysis with a Maurice instrument
- Control panel: Maurice Status - Communication with the instrument failed
- Message center: Run aborted - Instrument already in use / Maurice#xxxxx Instrument currently in use
- Reboot of the Maurice and the LAC/E does not fix the issue
ENVIRONMENT
- Empower 3 Enterprise
- Maurice driver from proteinsimple
CAUSE
Unknown
FIX or WORKAROUND
- Install the BioTechne software (Compass for iCE).
-
Using this software, clear the connection error.
ADDITIONAL INFORMATION
In the specific case that prompted this article, a cartridge purge (mandatory after disconnection) was done using Compass for iCE.
The instrument was then disconnected from Compass and reconnected to Empower without further problems.
(Normally, cartridge purge can be done under Empower, but not in this case related to disconnection.)
https://www.bio-techne.com/resources...s-software-ice