Discrepancy in the Empower 3.8.1 Installation, Configuration, and Upgrade Guide related to DCOM settings - WKB296368
SYMPTOMS
- Discrepancy in the Empower 3.8.1 Installation, Configuration, and Upgrade Guide (ICUG) related to DCOM settings
ENVIRONMENT
- Empower 3.8.1
CAUSE
Documentation error in the Empower 3.8.1 Installation, Configuration, and Upgrade Guide (ICUG)
FIX or WORKAROUND
Review the information in the additional information section of this Knowledge Base article.
ADDITIONAL INFORMATION
The Empower 3.8.1 Installation, Configuration, and Upgrade Guide (ICUG) contains an error in the DCOM settings and permissions set during installation table in sections 4, 5, 6, 8, and 9.
The path is Component Services > Computers > My
Computer > DCOM Config > WatersService >Properties > Security.
The Empower 3.8.1 ICUG currently states:
Under Launch and Activation Permissions,
set Local Launch, Local Activation, and
Remote Activation Permissions for:
• Everyone
• System
• Administrators
• Interactive
• Domain Users
The Empower 3.8.1 ICUG should state (because this is what the Empower installer sets):
Under Launch and Activation Permissions,
set Local Launch, Local Activation, Remote Launch, and
Remote Activation Permissions for:
• Everyone
• System
• Administrators
• Interactive
• Domain Users
Additionally, there are inconsistencies in the path: Component Services > Computers > My
Computer > DCOM Config > WatersService > Properties > Security
The Empower 3.8.1 ICUG states:
Under Access Permissions, set Local
Access and Remote Access Permissions
for:
• Everyone
• SELF
• System
• Administrators
• Domain Users
What is actually set by the Empower 3.8.1 installer is:
Under Access Permissions, set Local
Access and Remote Access Permissions
for:
• SELF
• System
• Administrators
***It is necessary to manually add the Everyone group.