Recovered injections and sample sets show corrupted timestamps in Empower with ORA-01877 - WKB294618

Article number: 294618

SYMPTOMS

  • Recovered injections have corrupted timestamps
  • Associated sample set finished timestamp may also be corrupted
  • Oracle Error "ORA-01877: string is too long for internal buffer" is recorded in the Empower Message Center
     
  • clipboard_e1637876f690eb0f064417496aa2fadfa.png

ENVIRONMENT

  • Linux Empower database
  • Timezone files of clients match the database server

CAUSE

This problem is likely due to intermittent network issues that occur when the time is being written – especially during recovery from buffering.

 

FIX or WORKAROUND

None. The timezone is corrupted and cannot be changed or updated.