"Data missing" is seen in the Review window of Empower 3, and ORA-01110 and ORA-01115 are seen in the message center - WKB29368

Article number: 29368

SYMPTOMS

  • ORA-01110
  • ORA-01115
  • "Data missing" is seen for all data in a particular project
  • Data can be reviewed as expected in other projects

ENVIRONMENT

  • Empower 3 Personal

CAUSE

Corrupted schema.

FIX or WORKAROUND

  1. Back up the project with this issue, and then restore it, giving it a different name.
  2. Confirm that you can open this project and review data.
  3. Back up the new project.
  4. Delete the original project.
  5. Restore the new project and give it the same name as the original project.

ADDITIONAL INFORMATION

 

 

 