The Empower 3 Deployment Manager has detected that Empower 3 software or any related process is running on this system. Please exit all applications and then run this Deployment Manager again - WKB27837
Article number: 27837
SYMPTOMS
- Error observed when upgrading or installing a driver: The Empower 3 Deployment Manager has detected that Empower 3 software or any related process is running on this system. Please exit all applications and then run this Deployment Manager again.
- Deployment Manager closes
ENVIRONMENT
- Empower 3
CAUSE
WatersEmailService is running.
FIX or WORKAROUND
- Go to Computer Management > Services and Applications > Services.
- Scroll down and select WatersEmailService.
- Click the link in the upper-left portion of the window to stop this service, or right-click and select Stop.
- Install Empower 3 FR4.