Empower error "It is illegal to call out while inside message filter" - WKB275249
Article number: 275249
SYMPTOMS
- System loses communications intermittently
- Displays "It is illegal to call out while inside message filter" error in Run Samples
- Requires LAC/E reboot to regain communications with instruments
ENVIRONMENT
- Empower Enterprise
- Empower Workgroup
- QDa
CAUSE
User Account Control is set to "Always Notify".
FIX or WORKAROUND
Reset User Account Control to "Never Notify" as described in the installation guide.
ADDITIONAL INFORMATION
Some customers have this policy set via group policy. If the error persists after resetting to Never Notify, check to ensure that it has not reverted back to Always Notify. You may need to work with customer IT or cybersecurity department to ensure that the setting does not revert.