Empower and Arc HPLC run sample sequence even when column heater/cooler is switched off - WKB268754
Article number: 268754
SYMPTOMS
- Column heater/cooler power turned off
- Empower still runs sequence to completion
- No faults in Empower message center
ENVIRONMENT
- Empower 3.7.0
- ACQUITY Arc
- ACQUITY Arc HPLC
- Arc HPLC 30-cm Column Heater/Cooler (part 186179100)
CAUSE
Unknown; CRI-6256 created for this issue
FIX or WORKAROUND
CRI-6256 was resolved in Driver Pack 2024 Release 1 (667007056DD).
ADDITIONAL INFORMATION
Set up the system to initial conditions before launching a sample set.
The Arc HPLC column heater is made from the same unit as the Alliance column heater/cooler for which this is a known issue. However, the workaround noted for use with the Alliance cannot be used with an Arc HPLC.