Empower and Alliance run sample sequence even when column heater/cooler is switched off - WKB242165

Article number: 242165

SYMPTOMS

  • Column heater/cooler turned off and gives a column temperature hw fault (7) on screen
  • Empower still runs sequence to completion

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module
  • 2795 Separations Module
  • Alliance Series Column Heater/Cooler

CAUSE

Known Alliance/Empower issue

FIX or WORKAROUND

On the Alliance console, choose develop methods > edit default, scroll to the column heater, and choose the stop function in place of disable.

ADDITIONAL INFORMATION

 