"Manual resolution required" error during method setup with Agilent LC - WKB262778
Article number: 262778
SYMPTOMS
- No changes to hardware or preconfiguration
- Method setup gives "System idle - instrument failure" message in the Empower Run Samples window
- Message center error shows "Method is inconsistent with current instrument configuration" error and Manual resolution required/Instrument failure error messages
ENVIRONMENT
- All Empower versions
- All ICF versions
- All Agilent LC instruments
CAUSE
Temperature control mode automatically changes from Variable to Constant temperature mode.
FIX or WORKAROUND
Right-click on sampler. Select mode > Temperature mode > Variable temperature mode.
ADDITIONAL INFORMATION
id262778,