Skip to main content
Waters

Intermittent instrument communication issues in Empower - WKB259688

Article number: 259688

SYMPTOMS

  • Intermittent instrument communication issues on Empower nodes
  • Instruments report "No" in the OK column of Instruments tab in Node Properties

ENVIRONMENT

  • Empower

CAUSE

Incorrect network topology

The switch used for the Instrument LAN is connected to the instrument LAN Ethernet card as expected. Additionally, there is an Ethernet connection to the Corporate network present on the same switch.

FIX or WORKAROUND

Remove the Ethernet connection from the Corporate network to the Ethernet switch supporting the Instrument LAN.

ADDITIONAL INFORMATION

 

id259688, comms, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.