Agilent LC system components are dimmed (unavailable) in the Empower Run Samples control panels from the Empower Citrix server but green (available) from the Empower LAC/E - WKB258263
Article number: 258263
SYMPTOMS
- Agilent column compartment and VWD modules are dimmed (unavailable) in the Empower Run Samples control panels from Citrix server
- Agilent column compartment and VWD modules show green (available) status in the Empower Run Samples control panels from LAC/E
ENVIRONMENT
- Instrument Control Framework Support Layer v3.1 on Empower Client
- Instrument Control Framework Support Layer v2.2 on Empower LAC/E
CAUSE
Incompatibility betweend the Instrument Control Framework Support Layer versions on the Empower client and Empower LAC/E
FIX or WORKAROUND
Upgrade the Instrument Control Framework Support Layer version on the LAC/E to v3.1.
ADDITIONAL INFORMATION
This behavior is documented in CRI-5717.
id258263, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP