Can chromatographic instruments and LAC/E be connected to a switch configured for VLAN? - WKB254649
Article number: 254649
ENVIRONMENT
- Empower
- LAC/E
- Chromatographic instruments
- Instrument LAN
ANSWER
No
This is not supported or tested.
- The Instrument LAN MUST be configured as prescribed in the Empower installation guide.
- It is expected that each LAC/E has a dedicated switch (if one is required to connect multiple instruments)
- Instrument should be connected directly to the instrument LAN of the LAC/E (this may use a dedicated switch)
ADDITIONAL INFORMATION
Waters cannot guarantee there will be no unforeseen issues connecting multiple LAC/Es and instruments to a switch using VLANs to segregate them.
- The impact on the buffering mechanism of the LAC/E is unknown and untested/unsupported
- Packet loss or other impacts on raw data file transfer, including header corruption
- More frequent need to verify incomplete data after acquisition completes
- Possibility that acquisitions will not run or be interrupted
The customer who asked wanted to mitigate cables being unplugged. This risk is not mitigated by using VLANs because the LAC/E or instrument connector can still be unplugged. There is no valid reason to employ VLAN for the instrument LAN.