Can chromatographic instruments and LAC/E be connected to a switch configured for VLAN? - WKB254649
Article number: 254649
ENVIRONMENT
- Empower
- LAC/E
- Chromatographic instruments
- Instrument LAN
ANSWER
No
This is not supported or tested.
- The Instrument LAN MUST be configured as prescribed in the Empower installation guide.
- It is expected that each LAC/E has a dedicated switch (if one is required to connect multiple instruments)
- Instrument should be connected directly to the instrument LAN of the LAC/E (this may use a dedicated switch)
ADDITIONAL INFORMATION
Waters cannot guarantee there will be no unforeseen issues connecting multiple LAC/Es and instruments to a switch using VLANs to segregate them.
- The impact on the buffering mechanism of the LAC/E is unknown and untested/unsupported
- Packet loss or other impacts on raw data file transfer, including header corruption
- More frequent need to verify incomplete data after acquisition completes
- Possibility that acquisitions will not run or be interrupted
The customer who asked wanted to mitigate cables being unplugged. This risk is not mitigated by using VLANs because the LAC/E or instrument connector can still be unplugged. There is no valid reason to employ VLAN for the instrument LAN.
id254649, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP