Selective shutdown method for pump flow is not working correctly in ICF SL 3.6 in Empower - WKB254653
Article number: 254653
SYMPTOMS
- The selective shutdown method is not working correctly in ICF SL 3.6 in Empower
- It may not be possible to put the pump into initial conditions, and it always moves to Standby
ENVIRONMENT
- Agilent Instrument Control Framework versions SL 3.5, SL 3.5.1, and SL 3.6
- Empower 3 Personal
- Empower 3 Workgroup
- Empower 3 Enterprise
CAUSE
In the Shutdown method, the Stop run was called in the ICF SL support layer instead of the Abort run.
FIX or WORKAROUND
CRI-4860 - Fixed in ICF Support Layer v3.7