Selective shutdown method for pump flow is not working correctly in ICF SL 3.6 in Empower - WKB254653

Article number: 254653

SYMPTOMS

  • The selective shutdown method is not working correctly in ICF SL 3.6 in Empower
  •  It may not be possible to put the pump into initial conditions, and it always moves to Standby

ENVIRONMENT

  • Agilent Instrument Control Framework versions SL 3.5, SL 3.5.1, and SL 3.6
  • Empower 3 Personal
  • Empower 3 Workgroup
  • Empower 3 Enterprise

CAUSE

In the Shutdown method, the Stop run was called in the ICF SL support layer instead of the Abort run.

FIX or WORKAROUND

CRI-4860 - Fixed in ICF Support Layer v3.7

ADDITIONAL INFORMATION

 