Cannot Save a Sample Set Method in Empower - WKB2531
Article number: 2531
SYMPTOMS
- Getting a Sample Set Method already in use error when trying to save a sample set method.
- Under Sample Set Method Properties, the Clear icon is unavailable, but "Being Edited By:" is populated.
- Tablespace error in the Message Center.
- In the Project Properties, the amount of Free Tablespace is low.
ENVIRONMENT
- Empower 3
CAUSE
Not enough Tablespace in the project properties.
FIX or WORKAROUND
- Go to Configure the System.
- Select Projects in the left-hand tree.
- Highlight the Project with the Tablespace error.
- Right-click and select Properties...
- Increase the Database Tablespace Quota, and then click OK.
- Close any windows with this project opened.
- Reopen Project and save Sample Set Method, and then run.
ADDITIONAL INFORMATION