Sample Set Name is missing from the Empower report - WKB2533
Article number: 2533
SYMPTOMS
- The report does not show the Sample Set Name
- Sample Set Names that have many characters do not show in the report
- Sample Set Names with few characters do show up in the report
ENVIRONMENT
- Empower 3
CAUSE
The cell for Sample Set Name in the header is not big enough to hold long names.
FIX or WORKAROUND
- Bring the Sample Set/Result Set into Preview/Publisher.
- Select the Report Method that has the issue.
- View the header/footer or, if you have the header on the first page, view the first page header.
- Ungroup the header.
- Manually make the size of the cell for Sample Set Name bigger.
- Preview the report to make sure that the size is sufficient.
- Regroup the header.
- Save the report method.
ADDITIONAL INFORMATION
The same holds true for the Sample Name field.