Error "Export Spawn Process failed for Project XXXX" during project restoration or backup - WKB251367

Article number: 251367

SYMPTOMS

  • Error while restoring or backing up project: "Export Spawn Process failed for Project XXXX. Check the Backup/Restore Wizard display for error messages for further details"

ENVIRONMENT

  • Empower (all versions)

CAUSE

Bad sector on hard disk. Oracle is unable to read/write the data file.

FIX or WORKAROUND

  1. Fix the hard disk issue.
  2. Refer to the Event Viewer to understand which hard disk is having an issue.