"Instrument Failure - failed at end of run" in Empower message center - WKB241411

Article number: 241411

SYMPTOMS

  • A running sample set is interrupted with "Instrument failure" in the Run Sample window and the following message in the Empower message center: "Instrument Failure - Wxxxx-xxxxxxxxx failed at end of run"

ENVIRONMENT

  • Empower 3 Enterprise

CAUSE

  • Possibly traffic congestion on an Instrument LAN bridge, or
  • The cable for communication with instrument modules is not properly shielded, or
  • Improper ground-to-neutral voltage (applicable only in certain regions)
  • 5V on low power supply

FIX or WORKAROUND

  1. Remove the instrument LAN bridge and use an external switch to connect the instrument modules. 
  2. Use a properly shielded CAT 6 cable for better communication.
  3. Ensure that the ground-to-neutral voltage is less than 2 V.
  4. Adjust 5V on low power supply to 5.2V on R-43 

ADDITIONAL INFORMATION

Check bios parameters : Empower 3 Instrument Failure - End of run wait timed out at 1800.8 seconds. No reply from SIIDetector#XXX - wkb283237

id241411, 2489, 2489B, 2489SFC, A-10UV, comms