Empower 3 "Instrument Failure - End of run wait timed out at 1800.8 seconds. No reply from SIIDetector#XXX" - WKB283237
Article number: 283237
SYMPTOMS
- Error occurs after a large sequence with Agilent + CAD instruments
- Instrument Failure SIIDetector#XXX
- There is an error in XML document (1, 1)
- Stopped the sequence queue run
- Instrument Failure - End of run wait timed out at 1800.8 seconds. No reply from SIIDetector#XXX
ENVIRONMENT
- Empower 3
- Agilent instrument
- SIIDetector
- LAC/E 15
- LAC/E 16
CAUSE
BIOS performance issue
FIX or WORKAROUND
- On a Dell-LAC/E Computer, disable C-State in the BIOS:
- Reboot PC.
- Press F2 while the computer is rebooting to enter the BIOS configuration menu.
- Under Settings, select Performance.
- Scroll down to "C-States Control".
- Clear the "C states" check box.
- Click the Apply button.
- Click the Exit button.