Empower 3 "Instrument Failure - End of run wait timed out at 1800.8 seconds. No reply from SIIDetector#XXX" - WKB283237

Article number: 283237

SYMPTOMS

  • Error occurs after a large sequence with Agilent + CAD instruments 
    • Instrument Failure SIIDetector#XXX
    • There is an error in XML document (1, 1)
    • Stopped the sequence queue run
    • Instrument Failure - End of run wait timed out at 1800.8 seconds. No reply from SIIDetector#XXX 

ENVIRONMENT

  • Empower 3
  • Agilent instrument
  • SIIDetector
  • LAC/E 15
  • LAC/E 16

CAUSE

BIOS performance issue

FIX or WORKAROUND

  1. On a Dell-LAC/E Computer, disable C-State in the BIOS:
    1. Reboot PC.
    2. Press F2 while the computer is rebooting to enter the BIOS configuration menu.
    3. Under Settings, select Performance.
    4. Scroll down to "C-States Control".
    5. Clear the "C states" check box.
    6. Click the Apply button.
    7. Click the Exit button.

ADDITIONAL INFORMATION