For the Empower 3 FR3 system audit trail, the "Archived and Removed Audit Trail" filter is not working (PCS 54126) - WKB235593

Article number: 235593

SYMPTOMS

  • Querying the System Audit Trail fails to list the expected entries when you apply the filter to bring all Archived and Removed records.
  • For the Empower 3 FR3 system audit trail, the "Archived and Removed Audit Trail" filter is not working

ENVIRONMENT

  • Empower 3 FR3

CAUSE

PCS 54126
PCS 56397

FIX or WORKAROUND

  1. The issue is fixed with Empower 3 SR3.
  2. You can create a view filter with this formula to list the event "Archived and Removed Audit Trail" :

    clipboard_e21131e3cad291ea812f839490fc8c413.png

    Details
    =*Removed By* *Records Before*

ADDITIONAL INFORMATION

 