ALS Error 945 for Agilent GC 7890B in Empower - WKB23480

Article number: 23480

SYMPTOMS

  • ALS Error 945 appears in Empower Message Center when trying to make an injection
  • Injection does not occur
  • Agilent 7890 shows Sampler error #945: Autoinjector back tower plunger error
  • Fault light on the tower flashes red four times

ENVIRONMENT

  • Empower 3 FR2 Enterprise
  • Agilent 7890 

CAUSE

Back injector plunger error. Faulty syringe or syringe carriage.

FIX or WORKAROUND

  1. Inspect the syringe and replace 
  2. If syringe replacement does not correct the error, replace the syringe carriage.  

ADDITIONAL INFORMATION

If it is a G4513A tower, order part G4513-67970 - RP Syringe Carriage. This is available to order through Waters. 