What does the Empower error "Data being acquired, problem reading data file" mean? - WKB234750

Article number: 234750

SYMPTOMS

  • Immediate error in the Review window when starting the run: "Data being acquired, problem reading data file"
  • Another error in the Review window: "No data acquired"
  • Message Center error example: "Lost connection to Agilent 7890A"

ENVIRONMENT

  • Empower 3 FR5 SR4 Enterprise
  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Lost Ethernet connectivity between the instrument LAN and the instrument

FIX or WORKAROUND

Reestablish connectivity.

ADDITIONAL INFORMATION

 