Agilent GC in dual tower mode waits until first injection is complete in Empower 3 - WKB233036

Article number: 233036

SYMPTOMS

  • GC waits until first injection is complete before second injection begins
  • Injections should occur simultaneously
  • Dual tower configuration per release notes of ICF Support Layer

ENVIRONMENT

  • Empower 3
  • Empower 3 Personal
  • Empower 3 Workgroup
  • Empower 3 Enterprise
  • Agilent Instrument Control Framework
  • Agilent 7890 GC

CAUSE

Sample Overlap is not enabled within Instrument Method.

FIX or WORKAROUND

Enable sample overlap under Tray in Instrument Method.

ADDITIONAL INFORMATION

 