"Invalid Vial ID Number" error message for an Agilent LC under Empower 3 control, but the sequence stops on the vial previous to the one cited in the message - WKB23220

Article number: 23220

SYMPTOMS

  • "Invalid Vial ID Number"
  • Run stops on previous vial
  • Previous vial is not acquired

ENVIRONMENT

  • Empower 3
  • Agilent Instrument Control Framework 2.1 (ICF v2.1)

CAUSE

The cause of the error in this case is that the user typed in vial number 111 when they meant to type 11. The reason that the injection prior to the vial labeled 111 is not collected is a defect (PCS 54156).

FIX or WORKAROUND

None

ADDITIONAL INFORMATION

id23220, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP