With Agilent LC, the sample set may stop after the buffering; the Run Sample window shows "Waiting for Post Run" when using Agilent ICFSL and Empower 3.6.1 - WKB229358
Article number: 229358
SYMPTOMS
- When using an Agilent LC chromatographic system on Empower 3.6.1, the sample set stops in "Waiting for Post Run" status and the sample set does not continue
- The issue arises only when the injections are recovering from buffering
ENVIRONMENT
- Empower 3.6.1
- Agilent ICF Support Layer (any version)
CAUSE
The post-time setting in the Instrument method is OFF, or the value is smaller than the time the system takes to recover the buffered data.
FIX or WORKAROUND
Add a non-zero post time value in any one of the instrument modules, or increase an existing post time value in the instrument method.
ADDITIONAL INFORMATION
This issue is reported as CRI-4380 which is resolved in Empower 3.7.0
For current Empower 3.6.1 customers, this issue is resolved in Hotfix, Empower 3.6.1.1