Serial port command error when using Agilent GC 6890N and Headspace 7694 in Empower - WKB22948
Article number: 22948
SYMPTOMS
- When run is started it stops, giving error message: HS7694@3 checksum error found in response to string HHSc0005A10- serial port command error
stuck at Vial 7 - Headspace communication okay
- Baud rate settings in 6890 are set up correctly
- Reboots of Headspace and GC do not resolve the issue
- Setup error can also cause instrument failure
ENVIRONMENT
- Empower 3 Feature Release 4
- Headspace 7694
- Agilent GC 6890N
- Edgeport
CAUSE
Corrupted method or methods
FIX or WORKAROUND
- Create new methods (instrument method and sample set method) in Empower.
- Rerun.
ADDITIONAL INFORMATION
- The 7694 is connected via RS232-to-USB adapter directly to LAC/E.
- The 6890N is connected via Edgeport (USB to serial converter)
-
See article WKB62833 on split injection technique.
See article WKB62831 on splitless injection technique.
See article WKB62933 on PTV injection technique.
id22948, 6890A/+GC, 6890N GC, comms, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, HEADSPACE, SUP