Serial port command error when using Agilent GC 6890N and Headspace 7694 in Empower - WKB22948

Article number: 22948

SYMPTOMS

  • When run is started it stops, giving error message: HS7694@3 checksum error found in response to string HHSc0005A10- serial port command error
    stuck at Vial 7
  • Headspace communication okay
  • Baud rate settings in 6890 are set up correctly
  • Reboots of Headspace and GC do not resolve the issue
  • Setup error can also cause instrument failure

ENVIRONMENT

  • Empower 3 Feature Release 4
  • Headspace 7694
  • Agilent GC 6890N
  • Edgeport 

CAUSE

Corrupted method or methods

FIX or WORKAROUND

  1. Create new methods (instrument method and sample set method) in Empower.
  2. Rerun.

ADDITIONAL INFORMATION

  • The 7694 is connected via RS232-to-USB adapter directly to LAC/E.
  • The 6890N is connected via Edgeport (USB to serial converter)
  • See article WKB62833 on split injection technique.

    See article WKB62831 on splitless injection technique.

    See article WKB62933 on PTV injection technique.

  • See article WKB97285 on Headspace Sample preparation