Method setup fails using Agilent GC 6890 in Empower 3 - WKB225804

Article number: 225804

SYMPTOMS

  • Message Center error: "Column 1 configuration method setup failed"
  • New instrument method and method set
  • Instrument will run 20-25 samples and stop

ENVIRONMENT

  • Empower 3 FR 4
  • Agilent 6890 GC

CAUSE

FID collector assembly is faulty.

FIX or WORKAROUND

Replace the FID collector assembly. 