The revision history for a sample set method is missing several versions when brought into Preview Publisher in Empower 3 - WKB22493

Article number: 22493

SYMPTOMS

  • When right-clicking the sample set method and selecting Preview/Publisher, the revision history is missing several method versions
  • The method versions summary table does show all versions; none are missing from this table

ENVIRONMENT

  • Empower 3

CAUSE

At some point, a version of the method was saved based on one of the earlier versions of the method. As an example, this may happen if there are currently five versions of a method and the sixth version is saved based on version 3 of the method. In this case, version 6 would have version 4 and 5 missing from the revision history when viewed in the Preview window. 

FIX or WORKAROUND

None. This scenario was described in PCS 55909, 56352, and 55284. As of Empower 3 Feature Release 4, this behavior still exists.

ADDITIONAL INFORMATION

 

 