Agilent PreConfiguration tool cannot connect to 1260 LC - WKB221822
Article number: 221822
SYMPTOMS
- Cannot use Auto Configure within the Agilent PreConfiguration tool
ENVIRONMENT
- Empower 3 Enterprise
- Empower 3
- Agilent 1260
- ICF Support Layer 3.5
CAUSE
The dip switches are incorrect.
FIX or WORKAROUND
-
On the back of the Agilent detector are dip switches.
The number of detector dip switches may be six or 8.eight.
To set up systems with eight dip switches, use:
1 2 3 4 5 6 7 8
Off Off Off Off On Off Off Off
To set up systems with six dip switches, use:
1 2 3 4 5 6
Off Off Off Off On Off
- Power cycle the Agilent detector—the stored IP address should be set.
ADDITIONAL INFORMATION
PreConfiguration Utility tool not working properly in Agilent LC - WKB958