Preconfiguration Utility tool not working properly in Agilent LC - WKB958
Article number: 958
SYMPTOMS
- Could not configure instrument using the Preconfiguration utility
ENVIRONMENT
- Empower Enterprise
- Agilent Instrument Control Framework (ICF) Support Version 2.2
CAUSE
User must type the Host (NIC Card) IP address first.
FIX or WORKAROUND
- Navigate to Empower > Configuration Manger > Tools > Agilent PreConfiguration.
- Either type the Host name or the IP address of the LAC/E NIC on the network (not the IP address of the instrument).
- Select New.
- Select Configure.
- Specify the IP address of the instrument.
- Configure as normal (as in previous version A.02.03).
ADDITIONAL INFORMATION
The Agilent.InstrumentControl.InstrumentPreConfigurator.exe changed in Agilent Instrument Control Framework (ICF) Support Version 2.2. It now can be accessed through Empower.
Agilent Instrument Control Framework (ICF) Support Version 2.2