UPLC CM error message “Power supply HW fault” - WKB95805

Article number: 95805

SYMPTOMS

  • CM error message “Power supply HW fault”
  • It is solved temporarily by reset, but the same error appears after a few minutes

ENVIRONMENT

  • ACQUITY UPLC Column Manager

CAUSE

700002623 (POWER SUPPLY, 420 WATT, 24VDC) failure

FIX or WORKAROUND

Replace the 700002623 (POWER SUPPLY, 420 WATT, 24VDC)

ADDITIONAL INFORMATION