Amounts do not generate for sample set in Empower 3 - WKB221176
Article number: 221176
SYMPTOMS
- Amounts populate for calibration curve
- Amounts do not populate for sample set run after curve
- Q20 and C02 peak codes
ENVIRONMENT
- Empower 3
CAUSE
Clear Calibration is the last line of the sample set.
FIX or WORKAROUND
Remove the final Clear Calibration line from the sample set.
ADDITIONAL INFORMATION
When using Clear Calibration, it should be the first line of a sample set.