Amounts do not generate for sample set in Empower 3 - WKB221176

Article number: 221176

SYMPTOMS

  • Amounts populate for calibration curve
  • Amounts do not populate for sample set run after curve
  • Q20 and C02 peak codes

ENVIRONMENT

  • Empower 3

CAUSE

Clear Calibration is the last line of the sample set.

FIX or WORKAROUND

Remove the final Clear Calibration line from the sample set.

ADDITIONAL INFORMATION

When using Clear Calibration, it should be the first line of a sample set.