Mn and Mz values are not being calculated for GPC data in Empower 3 - WKB21709

Article number: 21709

SYMPTOMS

  • The Mz and Mn values are not being calculated
  • Mw is calculated as expected
  • Injecting narrow standards and broad unknowns

ENVIRONMENT

  • Empower 3
  • Gel Phase Chromatography

CAUSE

The unknowns were injected as "unknowns" instead of "broad unknowns".

FIX or WORKAROUND

Take the samples into the Alter Sample window and change the sample type from unknown to broad unknown. Reprocess the data.

ADDITIONAL INFORMATION

 