Chromatograms do not appear in the Empower Review window - WKB209679
Article number: 209679
SYMPTOMS
- Chromatograms do not appear in the Empower Review window or on the client or LAC/E
- Chromatograms do not appear in the Empower Review window on the server
- Can ping the server from the LAC/E and client by IP address and vice versa
- Can ping the server from the LAC/E and client by name and vice versa, but very slow
ENVIRONMENT
- Empower Enterprise
CAUSE
The domain controller failed.
FIX or WORKAROUND
Fix the domain controller.
ADDITIONAL INFORMATION
id209679, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP