Agilent 1260 Infinity does not move to next injection in Empower - WKB209699
Article number: 209699
SYMPTOMS
- Agilent 1260 Infinity does not move to next injection in Empower
- Run time is set to 5 min and the first injection runs infinitely
- All modules meet the minimum firmware requirement for the ICF
ENVIRONMENT
- Agilent 1260 Infinity
- Empower Enterprise
- ICF 3.2
- G7111A
- G7129A
- G7130A
- G7114A
CAUSE
Preconfiguration has the correct setting but the instrument method has the wrong setting for the Thermostat in the ALS.
FIX or WORKAROUND
Set the Thermostat to Variable Temperature in the instrument method.
ADDITIONAL INFORMATION
id209699, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP