Agilent 1260 Infinity does not move to next injection in Empower - WKB209699

Article number: 209699

SYMPTOMS

  • Agilent 1260 Infinity does not move to next injection in Empower
  • Run time is set to 5 min and the first injection runs infinitely
  • All modules meet the minimum firmware requirement for the ICF 

ENVIRONMENT

  • Agilent 1260 Infinity
  • Empower Enterprise
  • ICF 3.2
  • G7111A
  • G7129A
  • G7130A
  • G7114A

CAUSE

Preconfiguration has the correct setting but the instrument method has the wrong setting for the Thermostat in the ALS.

FIX or WORKAROUND

Set the Thermostat to Variable Temperature in the instrument method.