Skip to main content
Waters

"Connection to chromatography system lost" with Empower - WKB208603

Article number: 208603

SYMPTOMS

  • "Connection to chromatography system lost" error
  • Acquisition stops

ENVIRONMENT

  • Empower 3
  • Empower 3 Personal
  • Empower 3 Workgroup
  • Empower 3 Enterprise

CAUSE

The acquisition computer or LAC/E32 went to sleep during run.

FIX or WORKAROUND

  1. On the acquisition computer or LAC/E32, navigate to the Windows Control Panel.
  2. Click "System and Security".
  3. Click "Power Options".
  4. On the Power Options page, click "Change Plan Settings".
  5. On the "Edit Plan Settings" page, change the following:
    1. Select "Never" from the "Turn off display" field.
    2. Select "Never" from the "Put the computer to sleep" field.
  6. On the "Edit Plan Settings" page, select "Change Advanced Power Settings".
  7. Edit the advanced power settings to match the following table:clipboard_ea3a81b6b3302027abb6c7865dd6df7c5.png

ADDITIONAL INFORMATION

Ensure that you check with your company IT before making any changes. Note that group policy may override these changes.

id208603, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.