No amounts are generated in Empower 3 when using default amounts - WKB20611
Article number: 20611
SYMPTOMS
- No amounts are generated in Empower 3 when using default amounts
ENVIRONMENT
- Empower 3
- Default amounts are being used
CAUSE
On the Default Amounts tab of the processing method, there is no entry for Level.
FIX or WORKAROUND
Specify a level for each injection of standard.
ADDITIONAL INFORMATION
When the Default Amounts tab is used, levels must be specified in order for a calibration curve to be created. This is different from use of the Component Editor to create a calibration curve.
This is also true in Breeze.