Unable to log in to LAC/E - WKB206173

Article number: 206173

SYMPTOMS

  • Error when trying to log in to domain and also local: "The number of connections to this computer is limited and all connections are in use right now. Try connecting later or contact your system administrator."
  • After clicking "OK" on the warning message, screen returns to login screen
  • Power cycling does not address this issue
  • It can still be seen from other LAC/Es or client PCs
  • Systems on this LAC/E can still run 
  • Regedit to increase Windows log in limits does not address issues

ENVIRONMENT

  • Empower 3 Enterprise

CAUSE

Unknown

FIX or WORKAROUND

Re-image the LAC/E.

ADDITIONAL INFORMATION

Multiple attempts to remote into this LAC/E simultaneously might have caused the sessions to glitch and hang up, and it's trying to reconnect to the the previous sessions on the LAC/E.