Agilent Pre-treatment functions from the instrument method are completely ignored during acquisition for all Agilent Samplers - WKB199396

Article number: 199396

SYMPTOMS

  • Pre-treatment functions like auto additions, dilutions, washes, and so on as configured in the instrument method are completely ignored during acquisition.

ENVIRONMENT

  • Agilent Instrument Control Framework v3.4 only
  • Empower 3 Enterprise
  • Agilent LC

CAUSE

Implementation of new functionality in Agilent Instrument Control Framework v3.4 broke the behavior of Pre-treatment functions

FIX or WORKAROUND

  1. CRI - 3386 created for this issue
  2. This issue is fixed in Agilent Instrument Control Framework v3.5 and it shall be downloaded here. Click Here to download ICF v3.5

ADDITIONAL INFORMATION

 

id199396,