The export method in Empower adds pound signs (#) in front of various lines in reports - WKB19812
Article number: 19812
SYMPTOMS
- When a report is exported, pound signs are added in front of various lines
- This does not occur in the preview/publisher
- This occurs only after a recent upgrade from Empower 3 Feature Release 2 to Feature Release 3
ENVIRONMENT
- Empower 3 Feature Release 3
- Empower 3 Feature Release 4
- Empower 3 Feature Release 5
CAUSE
Defect (PCS 56107 and 55510)
CRI-2505
FIX or WORKAROUND
- It has been shown that if the font size is less than 10 in the report table, this issue will occur.
- Exporting has been shown to work correctly when the tables in the report are moved off the page.
A report with the table on the page area in preview/publisher does not export correctly.
Exports to >
Moving the tables of the page area in preview/publisher does export correctly.
Exports to >
ADDITIONAL INFORMATION
The fix for PCS 56107 and 55510 requires updated code and is currently planned for Empower 4.
See the attached file for full workaround details.