"Aborted Sample Set" in Action and a blank or missing user name in Empower Project Audit Trails - WKB198065
Article number: 198065
SYMPTOMS
- Empower does not capture the username in the Audit Trail record for the cases below.
- During acquisition, if a user presses the "Stop Flow " button on the instrument
- During acquisition, the system flow is interrupted because of instrument errors like "Lost Prime" or "System over pressure"
- In the Audit Trails Action, it shows Aborted Sample Set and Reason: User Aborted After Injection #1 Line # XX Sample set Aborted
- When the instrument causes a failure, the user name is not recorded in the project audit trail.
ENVIRONMENT
- Empower 3.8.0 and later
CAUSE
This behavior is recorded as CRI-2839.
FIX or WORKAROUND
N/A. Empower is working as expected.
ADDITIONAL INFORMATION
- The Empower Project Audit Trail record does not show a username in the Username column when a user stops flow at the instrument. This is because Empower does not know what user stopped the flow.
- The Empower Project Audit Trail record does not show a username in the Username column when an instrument failure occurs. This is because the failure was not caused by a user.
- In both cases above, in Empower version 3.7.1 and lower, the Reason shown in the Details column of the Project Audit Trail shows "User Aborted".
- In both cases above, in Empower version 3.8.0 and higher, the Reason shown in the Details column of the Project Audit Trail shows "Aborted".
- In Empower 3.8.0 and higher, when an abort is performed by a user during acquisition within Empower, the Reason shown in the Details column of the Project Audit Trail shows "User Aborted".
Information regarding the change made in Empower 3.8.0, is documented in Empower 3.8.0 Release Notes, 715008193 Ver. 03 as CRI-2839.
id198065, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP