In Empower, names in report do not match Project window - WKB197480
Article number: 197480
SYMPTOMS
- Issue affects sample set names and method names in report method groups and plot legends
- The names may have missing characters or added characters; typically affects only one character and occurs when an unprintable control character or a special character is used in the name
ENVIRONMENT
- Empower 3
- Report method
CAUSE
A known defect, opened as CRI-2622
FIX or WORKAROUND
Issue was fixed in Empower 3.6.1.
ADDITIONAL INFORMATION
Also reported as CRI-261, CRI-738, PCS 54938, and PCS 56483
CRI-2622 - Names in Report do not Match Project Window | Waters
For more information about CRI-738, see the following:
CRI-738 - Character missing in Sample Set name in Report | Waters