Instrument failure before injecting with Empower and Agilent Fraction Collector G7159B Infinity II - WKB195724

Article number: 195724

SYMPTOMS

  • Instrument failure before injecting

ENVIRONMENT

  • Empower
  • Agilent Fraction Collector G7159B Infinity II

CAUSE

Incorrect collector setting

FIX or WORKAROUND

  1.  On the “General” tab of the instrument method, Set the correct values for the “Fraction Collection Options”.
    Syntax: P3-B-1  (P3 = Rack; B = Column; 1 = Raw) P1 is rear right.

    clipboard_eb728574e28fd198d11b043b4468f8654.png
  2. Even if collection is “Timed-Based”,  the “Peak Triggers” first column must be configured with a detector (default setting). 

    clipboard_e6f881054c377e2e1bd420777f7cb0a8a.png
     

ADDITIONAL INFORMATION

 