Instrument failure before injecting with Empower and Agilent Fraction Collector G7159B Infinity II - WKB195724
Article number: 195724
SYMPTOMS
- Instrument failure before injecting
ENVIRONMENT
- Empower
- Agilent Fraction Collector G7159B Infinity II
CAUSE
Incorrect collector setting
FIX or WORKAROUND
- On the “General” tab of the instrument method, Set the correct values for the “Fraction Collection Options”.
Syntax: P3-B-1 (P3 = Rack; B = Column; 1 = Raw) P1 is rear right.

- Even if collection is “Timed-Based”, the “Peak Triggers” first column must be configured with a detector (default setting).