Empower calibration curve is missing points - WKB195772
Article number: 195772
SYMPTOMS
- Empower calibration curve is missing points
- Amounts are entered
- No points shows up in calibration curve
ENVIRONMENT
- Empower 3
CAUSE
Missing level in the Sample Set
FIX or WORKAROUND
- Right-click on the sample set and select Alter Sample.
- Set the levels for each standard and reprocess the sample set.