Empower calibration curve is missing points - WKB195772

Article number: 195772

SYMPTOMS

  • Empower calibration curve is missing points
  • Amounts are entered
  • No points shows up in calibration curve

ENVIRONMENT

  • Empower 3

CAUSE

Missing level in the Sample Set

FIX or WORKAROUND

  1. Right-click on the sample set and select Alter Sample.
  2. Set the levels for each standard and reprocess the sample set.

ADDITIONAL INFORMATION

 