Is it recommended to use two different instruments on same LAC/E? - WKB195700

Article number: 195700

ENVIRONMENT

  • Waters UPLC
  • Waters Alliance
  • Agilent GC 6890/7890/8890/Intuvo 9000
  • Agilent LC

ANSWER

It is not recommended to use Agilent LC, Agilent GC, or UPLC on the same LAC/E. Waters recommends use of only one type of instrument per LAC/E.

ADDITIONAL INFORMATION

The number of systems that can be connected to one Lace depends on the type of Lace and the systems you are trying to connect. At a maximum you can connect 4 systems. If you have PDA detectors, it is recommended that only two are online at once. For Beckman Coulter (Sciex) Capillary electrophoreses (CE) detectors, it is recommended to only connect one CE device per Lace box in an Empower Enterprise network.