Sample set won't move past the first injection in Empower - WKB19473
Article number: 19473
SYMPTOMS
- Sample set won't move past the first injection in Empower
- After the first injection, the sample set won't continue
ENVIRONMENT
- Empower 3
- Empower 2
CAUSE
The modules listed in the instrument method do not match the modules configured for the system in Empower.
FIX or WORKAROUND
Ensure that the modules in the instrument method match the modules configured for the system in Empower.
ADDITIONAL INFORMATION
Using only one detector in systems with multiple detectors may require creation and use of multiple systems based on the instrument method.